North Carolina Central University


A nanoparticle amplification based quartz crystal microbalance DNA sensor for detection of Escherichia coli O157:H7.

Publication Type  Journal Article
Author  Mao X, Yang L, Su X, Li Y
Year of Publication  2006
Secondary Title  Biosens Bioelectron
Volume  21
Pagination  1178--1185
Date Published  Jan
Type of Work  article
Publication Language  eng
Key Words  Biosensing Techniques
Abstract  

A quartz crystal microbalance (QCM) DNA sensor, based on the nanoparticle amplification method, was developed for detection of Escherichia coli O157:H7. A thiolated single-stranded DNA (ssDNA) probe specific to E. coli O157:H7 eaeA gene was immobilized onto the QCM sensor surface through self-assembly. The hybridization was induced by exposing the ssDNA probe to the complementary target DNA, and resulted in the mass change and therefore frequency change of the QCM. Streptavidin conjugated Fe(3)O(4) nanoparticles (average diameter=145 nm) were used as ``mass enhancers to amplify the frequency change. Synthesized biotinylated oligonucleotides as well as E. coli O157:H7 eaeA gene fragments (151 bases) amplified using asymmetric PCR with biotin labeled primers were tested. As low as 10(-12)M synthesized oligonucleotides and 2.67 x 10(2) colony forming unit (CFU)/ml E. coli O157:H7 cells can be detected by the sensor. Linear correlation between frequency change and logarithmic number of bacterial cell concentration was found for E. coli O157:H7 from 2.67 x 10(2) to 2.67 x 10(6)CFU/ml.

Citation Key  177